63561-25, CELLOMETER DISPOSABLE COUNTING CHAMBER, PLASTIC BASE, 25/BOX
These Disposable Cell Counters are an enclosed chamber with two ports for sample introduction. The chamber contains precisely spaced lines in a grid pattern. There are two counting chambers per slide for replicates. The counting grid pattern is the Improved Neubauer, as in a common Hemacytometer. The consistent pattern design allows the standard cell counting procedure to be followed.
With a pipetor, 15 to 20 microlilter cell-containing sample is introduced into the counting chamber through one of the ports. Liquid spreads inside the chamber, replaces air, which escapes through the opposing port.
63561-25, CELLOMETER DISPOSABLE COUNTING CHAMBER, PLASTIC BASE, 25/BOX
These Disposable Cell Counters are an enclosed chamber with two ports for sample introduction. The chamber contains precisely spaced lines in a grid pattern. There are two counting chambers per slide for replicates. The counting grid pattern is the Improved Neubauer, as in a common Hemacytometer. The consistent pattern design allows the standard cell counting procedure to be followed.
With a pipetor, 15 to 20 microlilter cell-containing sample is introduced into the counting chamber through one of the ports. Liquid spreads inside the chamber, replaces air, which escapes through the opposing port.
The cell counting chamber is placed under a phase contrast optical microscope with a typical magnification of 100. The microscope is focused both onto the grid pattern and the cell particles. Standard protocols are followed for cell counting.
Let the Cellometer™ replace your Hemacytometers, which are traditionally used for cell counting
Benefits of the Cellometer™
| Step | Cellometer™ | Hemacytometer |
| 1 | Mix cell suspension | Mix cell suspension |
| 2 | Pipette 15 to 20 microliter into sample port | Carefully load with 10 microliter solution |
| 3 | Count cells in 4 corner squares (C_total) | Count cells in 4 corner squares (C_total) |
| 4 | Cell Concentration = Dilution x (C_total / 4.0 ) x10 4 | Cell Concentration = Dilution x (C_total / 4.0 ) x10 4 |
| 5 | Dispose used Cellometer | Wash and dry cover slide and base |
| CP2 Test 1 | CP2 Test 2 | CG2 | |
| # of test (Cellometer™) | 15 | 19 | 17 |
| # of test (Hemacytometer) | 8 | 17 | 17 |
| Mean (Cellometer™) | 1.78 x 105 | 1.83 x 105 | 1.87 x 105 |
| Mean (Hemacytometer) | 1.85 x 105 | 1.82 x 105 | 1.82 x 105 |
| p-value | 0.325 | 0.908 | 0.505 |
| Manufacturer | Electron Microscopy Sciences |
|---|---|
| Size / Rxns Number: | No |
| Protocol & Manual | No |
| Technical Tips | No |